Common Candela Laser Fault Codes

June 23, 2021

Stay in the loop

"*" indicates required fields

Name*
This field is for validation purposes and should be left unchanged.

Do you have a fault code on your Candela device? Click on the fault to go straight to the potential fix, or call Laseraid service on (02) 9011 5509.

For urgent service and repairs, contact us immediately. Our team has the shortest turnaround time in the industry, parts on hand and technicians available 7 days a week.

 

Fault 1.1

HP bubble circuit. Replace the cryogen and if the problem persists, contact Laseraid for servicing.

Fault 1.2

Canister bubble circuit. Replace the cryogen and if the problem persists, contact Laseraid for servicing.

Fault 3.1

Shutter fault. Turn off the laser for 30 seconds and then restart the system. If the problem persists, contact Laseraid for servicing.

Fault 4.2

Power Supply Communication. Turn off the laser for 30 seconds and then restart the system. If the problem persists, contact Laseraid for servicing.

Fault 5.1

Power Supply Tolerance. Turn off the laser for 30 seconds and then restart the system. If the problem persists, contact Laseraid for servicing.

Fault 5.2

Power Supply Charge. Turn off the laser for 30 seconds and then restart the system. If the problem persists, contact Laseraid for servicing.

Fault 5.4

Power Supply Charging Rate. Turn off the laser for 30 seconds and then restart the system. If the problem persists, contact Laseraid for servicing.

Fault 6.1

Calibration. Calibrate the laser system. If the problem persists, contact Laseraid for servicing.

 

Fault 6.5

Call port switch fault. Re-insert handpiece in calport. If the problem persists, contact Laseraid for servicing.

 

Fault 7.1

DI Water Low Temperature. Put the laser on Standby and allow sufficient time for the system to warm-up

Verify the laser room environment and temperature meet the specification requirement

Turn off the laser for 30 seconds and then restart the system.

If the problem persists, contact Laseraid for servicing.

 

Fault 7.2

Room Temperature may be too hot. Check the Aircon temperature on Aircon Control Panel. This should be set to no warmer than 18 degrees.

Has your device been moved recently? Perhaps it is much closer to a wall now than before and there may not be enough ventilation. Make sure there is sufficient space between your device and the walls of your treatment room.

Fault 7.3

Check your device’s water tank and the level of the water in the tank. The water level should be just under the lip, top-up water takes to this level if it is a little low. Your machine MUST cool down before water is added. Turn the machine off for 30 minutes prior to adding water.

Cause –  the change in temperature shocks the glass parts inside your machine and can cause them to break or be damaged.

Use distilled water to refill ONLY not demineralised.

Fault 8.1

Cryogen counter on ‘0’? You should see a number not ‘0’.*

Reset the counter by pressing and holding down the cryogen counter button on your screen.

If the symptom persists:

Take out the canister, shake it to see if it has gas in it or not. – replace with a new one. Allow to warm up and try again. Call Laseraid after 20 min if fault persists.

*Keep in mind that the counter does not measure the amount of cryogen in the canister.

Fault 8.2

Cryogen Canister may have been heated at a rapid rate causing excess pressure within the canister and causing this fault. Replace the canister and allow the device to heat up the canister and try again.

 

Fault 10.1

Remove the handpiece from the device and reconnect it, set spot size, calibrate, and try again.

 

Fault 10.4

Remove fibre from the machine and re-insert it, making sure there is nothing in the way or stopping it from going all the way in. You may need to do this a couple of times.

Calibrate & Pulse, this should clear your fault.

Fault 10.5

Your slider button on your handpiece may be broken. Test this by gently wiggling the fibre back and forth to see if it is loose or holding in place.

If it is loose, your handpiece will need to be repaired to clear this fault.

 

Fault 13.1

Your trigger button may not have been pressed directly in the middle.

Press trigger firmly in the middle to see if it is happening again.

Fault 15.1

Put the machine into ‘Standby.’

Remove your handpiece completely from the machine, point the end of the fibre that inserts into your device to the ceiling light and look into the handpiece, can you see the light through the hand piece? – no? it is possible that your fibre is broken and will need to be replaced.

Remove the window from the handpiece, point fibre to the ceiling light and take another look into the handpiece, can you see and dark spots, discolouration or burn marks on the internal window/ lenses? These will need to be replaced to clear the fault.

 

Fault 15.2

It is possible that the incorrect window may have been replaced when preparing your device for your treatments. Remove the window from the handpiece & try pulsing again.

If fault clears use a new window and try again.

 

Exit to clean window / Low Transmission

Mini Series devices – Replace the window located at the front of your slider. Ensure the window is unused and does not have any markings, dust etc when replace.  Once done recalibrate and try again.

Pro Series devices – Replace the window located at the front of your slider as well as the window located within the window draw in your hand piece. Again, ensure the window is unused and does not have any markings, dust etc when replaced.  Once done recalibrate and try again.

 

Replace Canister

Replace DCD canister & reset the canister counter whenever this message appears.

 

Purge

Air was detected in the cryogen lines, with handpiece removed from the calibration port, depress the purge button until the message clears.

 